Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and mu...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2665-2014.
Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management System, software version 14. The issue may cause the software to stop executing commands, including uploading validated results to the LIS. The issue is related to an internal software timer that overflows after 24 days. Automated rules triggered by task schedulers or by actions in the CentraLink user interface may also be affected. As each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2665-2014.
Recall terminated by FDA.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2665-2014.
Siemens Healthcare Diagnostics, Inc.
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Siemens Healthcare Diagnostics, Inc. Recall FAQ
Siemens Healthcare Diagnostics, Inc. is the subject of a medical devices safety report: Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and mu.... The notice was published on August 19, 2014 by the U.S. Food and Drug Administration (FDA). Approximately 858 units are potentially affected.