The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, ...
⚠ Critical FDA Safety Notice — Talk to Your Doctor
The FDA has flagged this as a serious safety concern (fire or burn). Do not stop, change, or discard a prescribed medication or medical device on your own. Contact your doctor or pharmacist immediately to ask whether this notice applies to your prescription and what to do next. Any specific guidance below comes directly from the FDA — Product Recall Tracker does not provide medical advice.
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2476-2021.
Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2476-2021.
Recall terminated by FDA.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2476-2021.
Philips North America Llc
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Philips North America Llc Recall FAQ
Philips North America Llc is the subject of a cardiac devices safety report: The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, .... The notice was published on August 13, 2021 by the U.S. Food and Drug Administration (FDA). Approximately 1 units are potentially affected.