CriticalFDAfda-Z-2073-2019ELECTRICAL SHOCK

ELLIPSE DR, REF: CD2377-36QC Implantable cardioverter defibrillators (ICDs)

⚠ Critical Alert — Stop Using Immediately

This product has been flagged with severe risks (electrical shock). Stop using it now and contact the brand or FDA for a refund, repair, or replacement.

Units Affected
21
Recall Date
June 20, 2019
Issuing Agency
Hazard
Electrical Shock

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2073-2019.

Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2073-2019.

Recall terminated by FDA.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact St Jude Medical Inc. or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2073-2019.

St Jude Medical Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

St Jude Medical Inc. Recall FAQ

St Jude Medical Inc. is the subject of a medical implants safety report: ELLIPSE DR, REF: CD2377-36QC Implantable cardioverter defibrillators (ICDs). The notice was published on June 20, 2019 by the U.S. Food and Drug Administration (FDA). Approximately 21 units are potentially affected.