Skip to main content
CriticalFDAfda-Z-1947-2026CHEMICAL EXPOSURE

AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the ...

⚠ Critical FDA Safety Notice — Talk to Your Doctor

The FDA has flagged this as a serious safety concern (chemical exposure). Do not stop, change, or discard a prescribed medication or medical device on your own. Contact your doctor or pharmacist immediately to ask whether this notice applies to your prescription and what to do next. Any specific guidance below comes directly from the FDA — Product Recall Tracker does not provide medical advice.

Units Affected
21
Recall Date
March 27, 2026
Issuing Agency
Hazard
Chemical Exposure

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1947-2026.

Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1947-2026.

Recall ongoing. Follow firm instructions.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1947-2026.

North American Rescue LLC.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

North American Rescue Llc. Recall FAQ

North American Rescue Llc. is the subject of a diabetes devices safety report: AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the .... The notice was published on March 27, 2026 by the U.S. Food and Drug Administration (FDA). Approximately 21 units are potentially affected.