Skip to main content
CriticalFDAfda-Z-1797-2026SERIOUS INJURY OR DEATH

Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7

⚠ Critical FDA Safety Notice — Talk to Your Doctor

The FDA has flagged this as a serious safety concern (serious injury or death). Do not stop, change, or discard a prescribed medication or medical device on your own. Contact your doctor or pharmacist immediately to ask whether this notice applies to your prescription and what to do next. Any specific guidance below comes directly from the FDA — Product Recall Tracker does not provide medical advice.

Units Affected
1,240,115
Recall Date
March 12, 2026
Issuing Agency
Hazard
Serious Injury Or Death

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1797-2026.

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1797-2026.

Recall ongoing. Follow firm instructions.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1797-2026.

Insulet Corporation

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Insulet Corporation Recall FAQ

Insulet Corporation is the subject of a medical devices safety report: Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7. The notice was published on March 12, 2026 by the U.S. Food and Drug Administration (FDA). Approximately 1,240,115 units are potentially affected.