Skip to main content
High RiskFDAfda-Z-1727-2014PRODUCT QUALITY

Alere Triage¿ D-Dimer Test PN 98100, Lot W53884B. The Alere Triage¿ D-Dimer Test is a fluorescence immunoassay to be used with the Aler...

Units Affected
373
Recall Date
May 2, 2014
Issuing Agency
Hazard
Product Quality

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1727-2014.

Alere initiated this recall because a limited number of Alere Triage¿ D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sample these affected devices will either generate an error code (e.g. E4 error), or they may generate an incorrect result.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1727-2014.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1727-2014.

Alere San Diego, Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Alere San Diego, Inc. Recall FAQ

Alere San Diego, Inc. is the subject of a diagnostic devices safety report: Alere Triage¿ D-Dimer Test PN 98100, Lot W53884B. The Alere Triage¿ D-Dimer Test is a fluorescence immunoassay to be used with the Aler.... The notice was published on May 2, 2014 by the U.S. Food and Drug Administration (FDA). Approximately 373 units are potentially affected.