Artis systems - Product Usage: an angiography systems developed for single and biplane diagnostic imaging and interventional procedures i...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1496-2021.
Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the belts which may result in limited functionality of the Artis system up to system failure. Unintended movement of the C-Arm may cause a hazardous situation to the patient, operator, or staff members and may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1496-2021.
Recall terminated by FDA.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1496-2021.
Siemens Medical Solutions USA, Inc
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Siemens Medical Solutions Usa, Inc Recall FAQ
Siemens Medical Solutions Usa, Inc is the subject of a diagnostic devices safety report: Artis systems - Product Usage: an angiography systems developed for single and biplane diagnostic imaging and interventional procedures i.... The notice was published on March 30, 2021 by the U.S. Food and Drug Administration (FDA). Approximately 3,320 units are potentially affected.