Skip to main content
CriticalFDAfda-Z-1383-2021FIRE OR BURN

Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT

⚠ Critical FDA Safety Notice — Talk to Your Doctor

The FDA has flagged this as a serious safety concern (fire or burn). Do not stop, change, or discard a prescribed medication or medical device on your own. Contact your doctor or pharmacist immediately to ask whether this notice applies to your prescription and what to do next. Any specific guidance below comes directly from the FDA — Product Recall Tracker does not provide medical advice.

Units Affected
5,928
Recall Date
March 4, 2021
Issuing Agency
Hazard
Fire Or Burn

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1383-2021.

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1383-2021.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1383-2021.

Covidien, LP

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Covidien, Lp Recall FAQ

Covidien, Lp is the subject of a medical devices safety report: Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT. The notice was published on March 4, 2021 by the U.S. Food and Drug Administration (FDA). Approximately 5,928 units are potentially affected.