Allura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventiona...

Units Affected
2
Recall Date
December 1, 2016
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0924-2017.

Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0924-2017.

Recall terminated by FDA.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact Philips Electronics North America Corporation or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0924-2017.

Philips Electronics North America Corporation

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

Philips Electronics North America Corporation Recall FAQ

Philips Electronics North America Corporation is the subject of a medical implants safety report: Allura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventiona.... The notice was published on December 1, 2016 by the U.S. Food and Drug Administration (FDA). Approximately 2 units are potentially affected.