Skip to main content
High RiskFDAfda-Z-0771-2019OTHER

Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse Treatment Planning Syst...

Units Affected
1,450
Recall Date
December 26, 2018
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0771-2019.

There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) XY R >R (R is too big).

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0771-2019.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0771-2019.

Varian Medical Systems, Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Varian Medical Systems, Inc. Recall FAQ

Varian Medical Systems, Inc. is the subject of a medical devices safety report: Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse Treatment Planning Syst.... The notice was published on December 26, 2018 by the U.S. Food and Drug Administration (FDA). Approximately 1,450 units are potentially affected.