High RiskFDAfda-Z-0717-2018TIPPING HAZARD

Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd fo...

Units Affected
1,486
Recall Date
December 20, 2017
Issuing Agency
Hazard
Tipping Hazard

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0717-2018.

Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0717-2018.

Recall terminated by FDA.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact Philips Electronics North America Corporation or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0717-2018.

Philips Electronics North America Corporation

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

Philips Electronics North America Corporation Recall FAQ

Philips Electronics North America Corporation is the subject of a medical implants safety report: Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd fo.... The notice was published on December 20, 2017 by the U.S. Food and Drug Administration (FDA). Approximately 1,486 units are potentially affected.