Dynasty Liners: REF DLCOGA28, REF DLCOGB32, REF DLCOGC36, REF DLCOGD38, REF DLCOGE40, REF DLCOGF42, REF DLCOGG44, REF DLCOGH48, REF DLCO...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0684-2017.
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0684-2017.
Recall terminated by FDA.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0684-2017.
MicroPort Orthopedics Inc.
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Microport Orthopedics Inc. Recall FAQ
Microport Orthopedics Inc. is the subject of a medical devices safety report: Dynasty Liners: REF DLCOGA28, REF DLCOGB32, REF DLCOGC36, REF DLCOGD38, REF DLCOGE40, REF DLCOGF42, REF DLCOGG44, REF DLCOGH48, REF DLCO.... The notice was published on June 21, 2016 by the U.S. Food and Drug Administration (FDA). Approximately 6,250 units are potentially affected.