Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lacta...

Units Affected
13,435
Recall Date
October 11, 2018
Issuing Agency
Hazard
Crash

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0499-2019.

Siemens Healthcare Diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase (LDI) may exhibit an erroneously elevated result when the preceding assay in the reaction cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a component of the LA reagent, has been identified as the cause of the potentially elevated result. A falsely elevated result for lactate dehydrogenase may lead to additional investigations to determine the etiology of tissue damage. Results would be correlated with clinical history and presentation in addition to other diagnostic laboratory testing.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0499-2019.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0499-2019.

Siemens Healthcare Diagnostics, Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Siemens Healthcare Diagnostics, Inc. Recall FAQ

Siemens Healthcare Diagnostics, Inc. is the subject of a diagnostic devices safety report: Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lacta.... The notice was published on October 11, 2018 by the U.S. Food and Drug Administration (FDA). Approximately 13,435 units are potentially affected.