Skip to main content
High RiskFDAfda-Z-0400-2021CHEMICAL EXPOSURE

ARTIS Icono Interventional Fluoroscopic X-Ray system - Product Usage: Intended Use: ARTIS is a family of dedicated angiography systems de...

Units Affected
100
Recall Date
September 18, 2020
Issuing Agency
Hazard
Chemical Exposure

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0400-2021.

There is a software problem which affects the DSA Roadmap application on ARTIS Icono and ARTIS Pheno systems with software version VE20B. In rare cases, when the DSA acquisition has been started while the C-Arm or table is moving, there may be a shift between the image acquired at the start position and the image (i.e. DSA vessel map) taken at the reference position. This may lead to a DSA vessel map being overlaid to a subtracted fluoro image at a position that is not accurate.. However, if the displacement is slight or at an unfavorable plane, the user might rely on incorrect visualization of the catheter relative to the vessel map. This could imply danger to the patient.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0400-2021.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0400-2021.

Siemens Medical Solutions USA, Inc

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Siemens Medical Solutions Usa, Inc Recall FAQ

Siemens Medical Solutions Usa, Inc is the subject of a diagnostic devices safety report: ARTIS Icono Interventional Fluoroscopic X-Ray system - Product Usage: Intended Use: ARTIS is a family of dedicated angiography systems de.... The notice was published on September 18, 2020 by the U.S. Food and Drug Administration (FDA). Approximately 100 units are potentially affected.