ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43mm Item Number: 00-8065-54...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0323-2021.
Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0323-2021.
Recall terminated by FDA.
✅ What you should do
- Stop using the product if you own it.
- Check the model number, lot code, or sell-by date against the recall notice above.
- Contact Zimmer Biomet, Inc. or the retailer where you bought it for a refund, replacement, or repair.
- For the most current official instructions, visit the FDA recall page.
- If you've been hurt by this product, report the incident to FDA.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0323-2021.
Zimmer Biomet, Inc.
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report a food, supplement, or cosmetic problem to the FDA
If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.
Zimmer Biomet, Inc. Recall FAQ
Zimmer Biomet, Inc. is the subject of a medical implants safety report: ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 43mm Item Number: 00-8065-54.... The notice was published on September 24, 2020 by the U.S. Food and Drug Administration (FDA). Approximately 412 units are potentially affected.