Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0041-2024.
CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0041-2024.
Recall ongoing. Follow firm instructions.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0041-2024.
Blue Belt Technologies, Inc
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Blue Belt Technologies, Inc Recall FAQ
Blue Belt Technologies, Inc is the subject of a medical devices safety report: Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299. The notice was published on September 7, 2023 by the U.S. Food and Drug Administration (FDA). Approximately 56 units are potentially affected.